Description
CISPR 15 edition 9 offers different methods for the measurement of radio disturbance characteristics of electrical lighting equipment. One of these is the CDNE method. This method specifies the use of a coupling / decoupling network for emission measurement (CDNE) to measure disturbance voltages in the 
30 to 300 MHz frequency range. This method enables EUTs to be connected directly to the CDNE, allowing a single conducted emission measurement to replace a lengthy radiated emission test. 
The Teseq CDNEs are compliant with the actual versions of CSIPR 16-1-2, CISPR 16-2-1 and CISPR 15 edition 9.
Using a CDNE instead of CDN offers improved measurement reproducibility due to standard’s requirements for more restrictive limits of asymmetrical impedance, phase angle, symmetrical impedance and internal attenuation. 
CISPR 15 edition 9 requires the termination of the mains supply cable of the EUT with a CDNE positioned on the reference-ground plane for the OATS, SAC or FAR measurement method. The receiver port of the CDNE is terminated with a 50 Ω impedance.
| Parameter | Value | 
|---|---|
| Frequency range | 30 to 300 MHz  | 
| Power rating (EUT- and AE port) | AC max. voltage (line to line) for CDNE M210, CDNE M310: 520 V  | 
| Current max | 10 A  | 
| Mains sockets (EUT- and AE port) | 4 mm, safety banana  | 
| Mains sockets (EUT- and AE port) for CDNE M310-USJP | NEMA 5-15/IEC C14  | 
| Common mode impedance (EUT port) | 150 Ω +10/-20 Ω  | 
| Phase angle (EUT port) | 0° ±25°  | 
| Differential mode impedance (EUT port) | 100 Ω ±20 Ω  | 
| Common mode impedance (EUT port) | 150 Ω +10/-20 Ω  | 
| Transducer factor/Voltage division factor (EUT/RF port) incl. internal 10 dB attenuator: | 20 dB ±1.5 dB  | 